Easy to use 3 point crystal vacuum
stage
180 ° rotary safety shield covers
crystal during reading
Double safety switches on shutter
operation
Foot operated shutter switch
Micro grain solid carbide vacuum
chuck points
Fast, accurate and easy to read digital/analog
readout
Geiger-Muller tube attached directly
to goniometer base
Angular encoder accuracy is +/- 1
second of arc rotation
One year warranty on all parts
Solid state 30 kv 1.0 Ma power generator
Digital readout of high voltage
Double diffracted X-Ray beam for
high accuracy
|
The Shure Check Fixture
consists of these five components:
- Solid state X-ray machine
- 3-point Vacuum chuck
- Rotary Angular Encoder with digital display
- Geiger-Muller Tube
- Digital/Analog Ratemeter
The check fixture measures the angle of the quartz wafer or blank by using
x-ray diffraction. A wafer or blank is positioned on the 3-point vacuum chuck,
an x-ray beam strikes it at the calculated Bragg angle, and the beam is diffracted
by the quartz into the Geiger-Muller tube. Peaks are displayed on Shure's
Digital/Analog Ratemeter. The analog meter gives the operator a quick average
setting while the microprocessor controlled digital display gives an exact
readout.
The opposite end of the 3-point chuck is linked to a rotary angular
encoder which transmits the exact angle in degrees, minutes, and seconds
to the digital electronic readout.
The Shure single station X-Ray Machine is equipped with a modern
air cooled tube submerged in a dielectric oil. A continuous duty 30 kv solid
state power supply is standard.
Shure Mfg
9012 West 51St Terrace, Merriam, Ks 66203 USA
TEL: (913) 262-1776 FAX: (913) 262-3685
Email: shure@shuremfg.com |
|